ReliOn BP200 Owner's Manual
SAFETY INFORMATION
INTENDED USE
This monitor is designed to measure blood pressure and pulse rate of adults within the
range of the designated arm cuff, following the directions provided in this instruction
manual. The monitor is designed for home use. Read the Safety Information in this
instruction manual before using this monitor.
To assure the correct use of the product, basic safety measures should always be
followed including the warnings and cautions listed in this instruction manual.
OPERATING THE DEVICE
Contact your physician for specific information about your blood pressure.
Self-diagnosis and treatment using measured results may be dangerous. Follow the
instructions of your physician or licensed healthcare provider.
DO NOT adjust medication based on measurement results from this blood pressure
monitor. Take medication as prescribed by your physician. Only a physician is
qualified to diagnose and treat High Blood Pressure.
The monitor is not intended to be a diagnostic device.
If battery fluid should get in your eyes, immediately rinse with plenty of clean
water. Contact a physician immediately.
Read all of the information in the instruction manual and any other literature in the
box before operating the unit.
Do not use this device on infants or persons who cannot express their intentions.
Read the Special Conditions section of the instruction manual if your systolic
pressure is known to be more than 220 mmHg. Inflating to a higher pressure than
necessary may result in bruising where the cuff is applied.
Operate the device only as intended. Do not use the device for any other purpose.
Dispose of the device, components and optional accessories according to applicable
local regulations. Unlawful disposal may cause environmental pollution.
Do not use a cellular phone near the device. It may result in an operational failure.
Use only authorized parts and accessories. Parts and accessories not approved for
use with the device may damage the unit.
Use only 1.5V Alkaline batteries with this device. Other types of batteries may
damage the unit.
Do not use in a location with moisture, or a location where water may splash on the
unit.
CARE AND MAINTENANCE
Do not subject the monitor to strong shocks, such as dropping the unit on the floor.
Do not submerge the device or any of the components in water.
Store the device and the components in a clean, safe location.
Changes or modification not approved by the manufacturer will void the user
warranty. Do not disassemble or attempt to repair the unit or components.
SAFETY SYMBOLS USED IN THIS INSTRUCTION MANUAL
WARNING
Indicates a potentially hazardous situation which, if not avoided,
could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation which, if not avoided,
may result in minor or moderate injury to the user or patient or
damage to the equipment or other property.
INSTRUCTION MANUAL
Automatic
Blood Pressure Monitor
Model HEM-741CREL
INTRODUCTION
Thank you for purchasing the ReliOn® HEM-741CREL Automatic Blood Pressure
Monitor.
Your new blood pressure monitor uses the oscillometric method of blood pressure
measurement. This means the monitor detects your blood’s movement through your
brachial artery and converts the movements into a digital reading. An oscillometric
monitor does not need a stethoscope so the monitor is simple to use.
The HEM-741CREL comes with the following components:
• Monitor • Wide Range Cuff • 4 “AA” Alkaline Batteries
• Instruction Manual printed in English and Spanish • Quick Start Guide
Please read this instruction manual thoroughly before using the unit.
Please keep for future reference.
For specific information about your own blood pressure, CONSULT YOUR
DOCTOR.
Fill in for future reference.
DATE PURCHASED: __________________
SERIAL NUMBER: ____________________
• Staple your purchase receipt here
SAVE THESE INSTRUCTIONS
5339081-6B
LIMITED WARRANTY
Your ReliOn® HEM-741CREL Automatic Blood Pressure Monitor, excluding the arm
cuff and batteries, is warranted to be free from defects in materials and workmanship
appearing within 5 years from the date of purchase, when used in accordance with the
instructions provided with the monitor. The arm cuff is warranted to be free from
defects in materials and workmanship appearing within one year from the date of
purchase when the monitor is used in accordance with the instructions provided with
the monitor. The above warranties extend only to the original retail purchaser.
We will, at our option, repair or replace without charge any monitor or arm cuff
covered by the above warranties. Repair or replacement is our only responsibility and
your only remedy under the above warranties.
To obtain warranty service contact Customer Service by calling 1-800-634-4350 for the
address of Inspection center.
Enclose the Proof of Purchase. Include a letter, with your name, address, phone
number, and description of the specific problem. Pack the product carefully to prevent
damage in transit. Because of possible loss in transit, we recommend insuring the
product with return receipt requested.
THE FOREGOING IS THE SOLE WARRANTY PROVIDED BY OMRON IN
CONNECTION WITH THIS PRODUCT, AND OMRON HEREBY DISCLAIMS
ANY OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING IMPLIED
WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR
PURPOSE. IMPLIED WARRANTIES AND OTHER TERMS THAT MAY BE
IMPOSED BY LAW, IF ANY, ARE LIMITED IN DURATION TO THE PERIOD
OF THE ABOVE EXPRESS WARRANTY.
OMRON SHALL NOT BE LIABLE FOR LOSS OF USE OR ANY OTHER
SPECIAL, INCIDENTAL, CONSEQUENTIAL OR INDIRECT COSTS,
EXPENSES OR DAMAGES.
This warranty provides you with specific legal rights, and you may have other rights
that vary by jurisdiction. Because of special local requirements, some of the above
limitations and exclusions may not apply to you.
FOR CUSTOMER SERVICE
Visit our web site at: www.omronhealthcare.com
Call toll free: 1-800-634-4350
Note:
POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for U.S.A. only)
This product has been tested and found to comply with the limits for a Class B digital
device, pursuant to part 15 of the FCC rules.
These limits are designed to provide reasonable protection against harmful interference
in a residential installation. The product generates, uses, and can radiate radio
frequency energy and, if not installed and used in accordance with the instructions,
may cause harmful interference to radio communications. However, there is no
guarantee that interference will not occur in a particular installation. If the product does
cause harmful interference to radio or television reception, which can be determined by
turning the product on and off, the user is encouraged to try to correct the interference
by one or more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the product and the receiver.
• Connect the product into an outlet on a circuit different from that to which the
receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for Canada only)
This digital apparatus does not exceed the Class B limits for radio noise emissions
from digital apparatus as set out in the interference-causing equipment standard entitled
“Digital Apparatus”, ICES-003 of the Canadian Department of Communications.
Cet appareil numérique respecte les limites de bruits radioeléctriques applicables aux
appareils numériques de Clase B prescrites dans la norme sur le materiel brouilleur:
“Appareils Numériques”, ICES-003 édictée par le minister des communications.
Changes or modifications not expressly approved by the party responsible for
compliance could void the user’s authority to operate the equipment.
SPECIFICATIONS
Model:
Display:
Measurement Range:
Accuracy/Calibration:
Inflation:
Deflation:
Rapid Air Release:
Measurement Method:
Power Source:
Battery Life:
Operating Temperature
/Humidity:
Storage Temperature
/Humidity/Air Pressure:
Main Unit Weight:
Main Unit Dimensions:
Cuff Size:
Cuff Circumference:
Memory:
Contents:
Optional Accessories:
HEM-741CRELN3 (HEM-8720-WM)
LCD Digital Display
Pressure: 0 to 299 mmHg, Pulse: 40 to 180/min.
Pressure: ±3mmHg or 2% of reading
Pulse: ±5% of reading
Automatic by electric pump
Automatic pressure release valve
Automatic exhaust valve
Oscillometric method
1.5V 4 "AA" batteries
Approx. 300 uses with 4 new alkaline batteries
50ºF to 104ºF (10ºC to 40ºC) /15 to 90% RH
-4ºF to 140ºF (-20ºC to 60ºC) /10 to 95% RH
700-1060 hPa
Approximately 11 1/4 oz. (319g)
not including batteries
Approx. 6"(l) × 4 7/8"(w) × 3 1/2" (h)
(152 mm × 123 mm × 88 mm)
Approx. 6"(w) × 22 1/8"(l) (151 mm × 563 mm)
Cuff tube 35 3/8" (900 mm)
Fits arm circumferences 9" to 17"
(220 mm to 420 mm)
Up to 30 readings per user
Main Unit, Wide Range Cuff, 4 "AA" Alkaline Batteries,
Instruction Manual, and Quick Start Guide
Wide Range Cuff 9'' - 17'' (220mm - 420mm)
NOTE: These specifications are subject to change without notice.
Manufactured for OMRON HEALTHCARE Co., Ltd.
Made in China
Distributed by OMRON HEALTHCARE, INC.
1925 West Field Court
Lake Forest, IL 60045 U.S.A.
www.omronhealthcare.com
© 2012 OMRON HEALTHCARE, INC.
ReliOn® is a registered trademark used under license by OMRON HEALTHCARE,
INC.
.................................................................
Distributed by:
Wal-Mart® Stores, Inc.
Bentonville, AR 72716
FCC STATEMENT
HEM-8720-WM_B_M02_120913.pdf
SETTING THE DATE AND TIME
WARNING
If battery fluid should get in your eyes, immediately rinse with
plenty of clean water. Contact a physician immediately.
CAUTION
Read all of the information in the instruction manual and any other
literature in the box before operating the unit.
CAUTION
This device is intended for use in measuring blood pressure and
pulse rate in the adult population. Do not use this device on infants
or persons who cannot express their intentions.
APPLYING THE CUFF ON THE LEFT ARM
1. Make sure the air plug is securely inserted
in the main unit.
2. Remove tight-fitting clothing from your upper
arm.
3. Apply the cuff to your left upper arm so
the arrow is centered on the inside of your arm
and aligned with your middle finger.
The air tube runs down the inside of your arm.
The bottom of the cuff should be
approximately 1/2 " above your elbow.
4. Wrap the cuff firmly in place around your
arm using the cloth fastening.
APPLYING THE CUFF ON THE RIGHT ARM
When taking a measurement using the right arm, use this instruction for Step 3.
Apply the cuff to your right upper arm so
the arrow and the air tube runs along the inside
of your arm. The bottom of the cuff should be
approximately 1/2 " above your elbow.
NOTE: Be careful not to rest your arm on the air tube. This will restrict the flow of air
to the cuff.
APPLYING THE ARM CUFF
1/2 inch (1 - 2 cm)
Air Tube
Arrow
Left Arm
Right Arm
Arrow
TAKING A MEASUREMENT
The monitor is designed to store the blood pressure and the pulse rate in the memory
for two people (USER A or USER B) every time a measurement is completed.
1. Sit on a chair with your feet flat on the floor.
Place your arm on a table so the cuff is level
with your heart.
Keep still and do not talk during measurement.
2. Select your USER ID (A or B).
3. Press the ON/OFF button. All symbols appear on the display.
The cuff starts to inflate automatically. As the cuff inflates, the monitor automatically
determines your ideal inflation level.
Remain still and do not move your arm until the entire measurement process is completed.
NOTE: To stop the inflation or measurement, push the ON/OFF button. The
monitor will stop inflating, start deflating, and will turn off.
4. Inflation stops automatically and the measurement is started.
As the cuff deflates, decreasing numbers appear on the display and the Heartbeat
Symbol ( ) flashes.
5. When the measurement is complete, the arm cuff completely
deflates. Your blood pressure and pulse rate are displayed.
The 1999 WHO/ISH guidelines for the management of hypertension are as follows:
This is a general guideline because blood pressure depends on age and morbidity.
A lower home blood pressure goal may be recommended for certain patients,
including diabetic patients, pregnant women, and patients with renal failure.
6. Press the ON/OFF button to turn the monitor off.
NOTE: The monitor will automatically turn
off after 2 minutes.
CARE AND MAINTENANCE
To keep your blood pressure monitor in the best condition and protect the unit from
damage, follow the directions listed below:
Do not forcefully bend the arm cuff or air tube. Do not fold tightly.
Clean the monitor with a soft dry cloth.
Do not use any abrasive or volatile cleaners.
Do not attempt to clean the cuff.
CAUTION
Do not submerge the device or any of the components in water. Do
not subject the monitor to extreme hot or cold temperatures,
humidity or direct sunlight.
CAUTION
Store the device and the components in a clean, safe location.
CAUTION
Do not subject the monitor to strong shocks, such as dropping the unit on the floor.
Remove the batteries if the unit will not be used for three months or longer. Always
replace all the batteries with new ones at the same time.
Use the unit in a manner consistent with the instructions provided in this manual.
CAUTION
Use only authorized parts and accessories. Parts and accessories not
approved for use with the device may damage the unit.
CAUTION
Changes or modification not approved by the manufacturer will void the user
warranty. Do not disassemble or attempt to repair the unit or components.
TROUBLESHOOTING TIPS
PROBLEM
No power.
No display appears on the unit.
Measurement values appear too
high or too low.
CAUSES AND SOLUTIONS
Replace all four batteries with new ones.
Check the battery installation for proper placement
of the battery polarities.
Blood pressure varies constantly. Many factors
including stress, time of day, how you wrap the
cuff, may affect your blood pressure.
Review the sections “Before Taking a
Measurement” and “Taking a Measurement”.
ERROR INDICATORS
1. Press the hook on the top of the battery cover and lift the
cover off in the direction of the arrow.
2. Install 4 “AA” size batteries so the + (positive) and
- (negative) polarities match the polarities of
the battery compartment as indicated.
3. Replace the battery cover.
BATTERY REPLACEMENT
Low Battery Indicator
When the Low Battery Indicator appears on the display, turn the monitor off and
remove all the batteries. Replace with four new batteries at the same time. Long-life
alkaline batteries are recommended.
NOTES: • Turn the unit off before replacing the batteries. If the batteries are removed
while the unit is still on, the date and time will be reset to that of the
previous use. However, the measurement results are not deleted.
• When the batteries are replaced, you may need to reset the date and time. If
“-:--” appears on the display, refer to “Setting the Date and Time”.
CAUTION
Dispose of the device, components and optional accessories according to
applicable local regulations. Unlawful disposal may cause environmental
pollution.
BATTERY INSTALLATION
UNIT DISPLAY
KNOW YOUR UNIT
Main Unit:
Components:
SET button
Wide Range Cuff
Air tube
Air plug
Quick Start
Guide
Instruction
Manual
4 "AA" Alkaline
Batteries
DISPLAY SYMBOLS
IRREGULAR HEARTBEAT SYMBOL ( )
When the monitor detects an irregular rhythm two or more times during the
measurement, the Irregular Heartbeat Symbol ( ) will appear on the display with the
measurement values.
An irregular heartbeat rhythm is defined as
a rhythm that varies by less than 25% from
the average rhythm or 25% more than the
average rhythm detected while the monitor
is measuring the systolic and diastolic
blood pressure.
If the irregular heartbeat symbol ( )
displays with your measurement results,
we recommend you consult your
physician. Follow the directions of your
doctor.
MOVEMENT ERROR SYMBOL ( )
The Movement Error Symbol ( ) is displayed if you move your body during the
measurement. Please remove the arm cuff, and wait 2-3 minutes. Reapply the arm cuff
and take another measurement.
AVERAGE VALUE SYMBOL ( )
The Average Value Symbol ( ) is displayed when you press the memory button.
The most recent average reading appears on the display screen.
HEARTBEAT SYMBOL ( )
The Heartbeat symbol flashes on the display at every heartbeat during the
measurement.
Normal Heartbeat
Irregular Heartbeat
Short Long
Blood
pressure
Pulse
Pulse
Blood
pressure
USING THE MEMORY FUNCTION
The monitor automatically stores up to 30 sets of measurement values (blood pressure
and pulse rate) for each user (A or B). When 30 sets of measurement values are stored,
the oldest record is deleted to save the most recent values.
AVERAGING FUNCTION
The monitor calculates the average reading based on the three most recent sets of
measurement values taken within 10 minutes of the most recent reading.
NOTES: • If two sets of measurement values are stored in the memory for the 10
minutes period, the average is based on the two sets of measurement
values. If one set of measurement values is stored, this is displayed as the
average.
• When viewing the reading taken without setting the date and time, "-:--" is
displayed instead of the date and time.
• The average reading will not be calculated without setting the date and
time.
TO DISPLAY THE MEASUREMENT VALUES
1. Select your USER ID (A or B).
2. Press the Memory button ( ).
The Average Symbol ( ) appears on the display when the average
reading is shown.
3. Press the Down button ( ) to display the most recent set of
measurement values on the screen. The values are displayed from the most
recent to the oldest.
NOTES:
• The Memory number appears for
a second before the pulse rate is
displayed. The newest set is
number “1”.
• The date and time are alternately
displayed with the measurement
values.
Press the Up ( ) or Down ( ) button
to display the next set of values.
NOTE:
Press and hold the Up or Down button to display the values faster.
TO DELETE ALL VALUES STORED IN THE MEMORY
The values stored in the memory are deleted by USER ID.
You cannot partially delete values stored in the memory. All values for
the user you select will be deleted.
1. Select your USER ID (A or B).
2. While holding the Memory button ( ), press the ON/OFF button down
for more than 2 seconds until the display appears as it does in the
following illustration.
3. Press the ON/OFF button to turn the monitor off.
NOTE: The monitor will automatically turn off after 2 minutes.
Set the monitor to the current date and time before taking a measurement for the first
time.
1. TO START THE SETTING
Press the SET button to adjust
the date and time.
The year flashes on the display.
2. SETTING THE YEAR
The year can be set between 2013 and 2040. When the
display reaches 2040, it will return to 2013.
Press the Up button ( ) to advance by one year.
Press the SET button to set the current year.
The month flashes on the display.
NOTES: • Press the Down button ( ) to decrease the
date and time values.
• Press and hold the button to increase or
decrease the date and time values faster.
3. SETTING THE MONTH
Press the Up button ( ) to advance by one month.
Press the SET button to set the current month.
The day flashes on the display.
4. SETTING THE DAY
Press the Up button ( ) to advance by one day.
Press the SET button to set the current day.
The hour flashes on the display.
5. SETTING THE HOUR
The time is set using AM or PM.
Press the Up button ( ) to advance by one hour.
Press the SET button to set the current hour.
The minute flashes on the display.
6. SETTING THE MINUTE
Press the Up button ( ) to advance by one minute.
Press the SET button to set the minute setting.
7. TO TURN THE MONITOR OFF
Press the ON/OFF button.
NOTE: The monitor will automatically turn off 2 minutes
after the minute has been set.
WARNING
Contact your physician for specific information about your blood pressure.
Self-diagnosis and treatment using measured results may be dangerous.
Follow the instructions of your physician or licensed healthcare provider.
WARNING
DO NOT adjust medication based on measurement results from this blood
pressure monitor. Take medication as prescribed by your physician. Only a
physician is qualified to diagnose and treat High Blood Pressure.
WARNING
The monitor is not intended to be a diagnostic device.
SPECIAL CONDITIONS
CAUTION
Inflating to a higher pressure than necessary may result in bruising where
the cuff is applied.
If your systolic pressure is known to be more than 220 mmHg, press and
hold the ON/OFF button until the monitor inflates 30 to 40 mmHg higher
than your expected systolic pressure, after the cuff starts to inflate.
NOTES: • Do not apply more pressure than necessary.
• The monitor will not inflate above 299 mmHg.
When the measurement is complete, the arm cuff completely deflates.
Your blood pressure and pulse rate are displayed.
Air plug disconnected.
Arm cuff not applied
correctly.
Air is leaking from the
arm cuff.
Cuff over or under
inflated.
Device error.
Batteries are worn.
Movement during
measurement.
Clothing is interfering
with the arm cuff.
Insert the plug securely. Refer to
“Applying the Arm Cuff”.
Apply the arm cuff correctly. Refer to
“Applying the Arm Cuff”.
Do not inflate the cuff above 299 mmHg.
Refer to “Taking a Measurement”.
Remove any clothing interfering with the
arm cuff. Refer to “Applying the Arm
Cuff”.
Replace the four batteries.
Refer to “Battery Installation”.
Repeat measurement. Remain still and do
not talk during measurement. Refer to
“Taking a Measurement”.
Repeat measurement. Remain still and do
not talk during measurement. Refer to
“Taking a Measurement”.
If “E2” appears repeatedly, inflate the cuff
manually until it is 30 to 40 mmHg above
your previous measurement result. Refer
to “Taking a Measurement”.
Replace the cuff with a new one.
Contact Customer Service.
The arm cuff was inflated
above 299 mmHg when
inflating the cuff manually.
SYMBOL
CAUSE
CORRECTION
ON/OFF button
Memory button
USER ID
Selection Switch
Up/Down ( ) buttons
Air Jack
Display
Battery compartment
: to display the older readings
: to display the recent readings
Systolic blood pressure
Diastolic blood pressure
Memory symbol
Average value symbol
Heartbeat symbol
Irregular heartbeat symbol
USER ID symbol (A/B)
Blood pressure level indicator
Pulse display and Memory number
Deflation symbol
Battery low indicator
Date/Time Display
Movement error symbol
Systolic blood pressure
HIGH
above 139 mmHg
PRE
130 - 139 mmHg
NORM
129 mmHg or less
Diastolic blood pressure
HIGH
above 89 mmHg
PRE
85 - 89 mmHg
NORM
84 mmHg or less
Alternating display
Battery Cover
BEFORE TAKING A MEASUREMENT
To ensure a reliable reading follow these recommendations:
1. Avoid eating, drinking alcohol, smoking, exercising, and bathing for 30 minutes
before taking a measurement. Rest for at least 15 minutes before taking the
measurement.
2. Stress raises blood pressure. Avoid taking measurements during stressful times.
3. Measurements should be taken in a quiet place.
4. Remove tight-fitting clothing from your left arm.
5. Sit on a chair with your feet flat on the floor. Rest your left arm on table so that the
cuff is at the same level as your heart.
6. Remain still and do not talk during the measurement.
7. Wait 2-3 minutes between measurements. The wait time allows the arteries to return
to the condition prior to taking the blood pressure measurement. You may need to
increase the wait time depending on your individual physiological characteristics.
8. Keep a record of your blood pressure and pulse readings for your physician. A
single measurement does not provide an accurate indication of your true blood
pressure. You need to take and record several readings over a period of time. Try to
measure your blood pressure at the same time each day for consistency.
Category / Blood pressure
Hypertension
High-Normal
Normal
Systolic blood pressure
above 139 mmHg
130 - 139 mmHg
129 mmHg or less
Diastolic blood pressure
above 89 mmHg
85 - 89 mmHg
84 mmHg or less
INSTRUCTION MANUAL
Automatic
Blood Pressure Monitor
Model HEM-741CREL
INTRODUCTION
Fill in for future reference.
DATE PURCHASED: __________________
SERIAL NUMBER: ____________________
5339081-6B
Thank you for purchasing the ReliOn® HEM-741CREL Automatic Blood Pressure
Monitor.
HEM-741CREL
Automatic
Blood Pressure
Monitor
Blood Pressure Monitor
Quick Start Guide
Related Manuals for ReliOn BP200
-
Health o meter Healthcare HDM575-18 Owner's Manual
2 pages
-
Health o meter Healthcare HDM770DQ1-05 Owner's Manual
2 pages
-
Health o meter Healthcare BFM920-01 Owner's Manual
2 pages
-
Health o meter Healthcare HDM520-18 Owner's Manual
2 pages
-
Health o meter Healthcare HDM076DQ-60 Owner's Manual
2 pages
-
Health o meter Healthcare HDM166DQ-60 Owner's Manual
2 pages
-
Health o meter Healthcare HDR743DQ1-41 Owner's Manual
2 pages
-
Health o meter Healthcare HDM037DQ-01 Owner's Manual
2 pages
-
Health o meter Healthcare HDC100KD-01 Owner's Manual
2 pages
-
Health o meter Healthcare HDM560DQ1-01 Owner's Manual
1 pages
-
ReliOn Healthcare BP100 Owner's Manual
32 pages
-
ReliOn Healthcare BP300 Owner's Manual
2 pages
-
ReliOn Healthcare BP200W Owner's Manual
49 pages
-
ReliOn Healthcare ReliOn Micro Owner's Manual
76 pages
-
ReliOn Healthcare ReliOn Confirm Owner's Manual
74 pages
-
ReliOn Healthcare ReliOn Ultima Owner's Manual
44 pages
-
ReliOn Healthcare ReliOn Prime Owner's Manual
70 pages
-
ReliOn Healthcare A1c Test Owner's Manual
2 pages
-
ReliOn Healthcare Ketone Test Strips Owner's Manual
2 pages
-
ReliOn Healthcare Temple Touch Thermometer Owner's Manual
2 pages
-
Omron Healthcare Healthcare BP760N (2014 Series) Owner's Manual
38 pages
-
Omron Healthcare Healthcare BP742N Owner's Manual
38 pages
-
Omron Healthcare Healthcare BP710N Owner's Manual
31 pages
-
Omron Healthcare Healthcare HJ-325 Owner's Manual
2 pages