HP Saw HP E4419B User's Manual | Page 98

HP Saw User's Manual - Saw HP E4419B.
Download
Page  of 204
4.8, 1501 votes
background image

HP4402.book : chapter2.fm    58  Fri Feb 12 15:32:45 1999

Power Meter Operation
Setting the Range

2-58

HP E4419B User’s Guide

Setting the Range

The power meter has no internal ranges which can be set. The only ranges
that can be set are those of the HP E-series power sensors. With an
HP E-series power sensor the range can be set either automatically or
manually. Use autoranging when you are not sure of the power level you
will be measuring. There are two manual settings either “LOWER” or
“UPPER”. The lower range is more sensitive than the higher range.
“LOWER” covers the range of -70 dBm to -13.5 dBm. “UPPER” covers the
range of -14.5 dBm to +20 dBm. The default is “AUTO”.

Note

The following procedure details how to set the range for channel
A. To set the range on channel B use the same procedure using the

 softkey.

To set the range:

1.

Press

,

 . The current setting of the

range is displayed under the

 softkey.

2.

To change this setting press

. A pop up window appears.

Use

 or

 to highlight your choice.

3.

To confirm your choice press

.

B Input Settings

System
Inputs

A Input Settings

Range

Range

Enter

0.581167s